CHAMBER: 23.0 °C ± 1.0 °C
HUMIDITY: 50.0 % RH ± 5%
Ambient Noise Floor < 18 dBA
NABL ISO/IEC 17025 ACCREDITED CLASS-1 ACOUSTIC METROLOGY DIVISION // CHENNAI PRIMARY LAB Traceable to NPL (India) & BIPM

Sound
CALIBRATION FACILITY

The MSIR INDIA Acoustic Metrology Facility provides NABL-accredited sound and noise instrumentation calibration in compliance with IEC 61672 and IEC 60942. Utilizing Brüel & Kjær pistonphones, multifrequency acoustic calibrators, and an ultra-low-noise anechoic chamber with wedge absorption, we calibrate sound level meters, noise dosimeters, and measurement microphones.

SOUND LEVEL RANGE

30 dB to 160 dB

SPL (Sound Pressure Level)

FREQUENCY RANGE

20 Hz to 20 kHz

Full Audio Spectrum

METER CLASS

Class 0, 1 & 2

IEC 61672-1 Compliant

ACOUSTIC CMC

± 0.15 dB

Precision Pistonphone

LIVE METROLOGY SIGNAL SIMULATOR & TELEMETRY STUDIO

Anechoic Acoustic Wave & 1/3 Octave Band Spectrum Simulator

CHANNEL 1 // SOUND MASTER REFERENCE BUS
ACQUISITION: REAL-TIME METROLOGY (120 FPS)

MEASURED TELEMETRY

94.02

EXPANDED UNCERTAINTY (k=2)

±0.12 dB

TRACEABILITY

NPL India

PARAMETER TUNING

Sound Level Output:94 dB
Thermal / Environmental Noise:
Signal Stream:
ISO/IEC 17025 Compliant Measurement

Evaluated as per GUM (JCGM 100:2008) guidelines with complete Type A and Type B expanded uncertainty analysis.

PRIMARY INSTRUMENTATION ARSENAL

Primary Reference Standards

Our sound capabilities are anchored by primary standards directly calibrated by NPL India and BIPM-recognized National Metrology Institutes.

Primary Acoustic StandardISO 17025

Brüel & Kjær Type 4228 Pistonphone

Precision Acoustic Pistonphone Standard

Generates an exact 124.0 dB SPL tone at 250 Hz based on precision piston displacement and barometer ambient air pressure correction.

Technical Highlights:

Sound Pressure Level:124.0 dB ± 0.09 dB
Frequency:250 Hz ± 1 %
Harmonic Distortion:< 1.5 %
DPLA / NPL Direct Acoustic Traceability
ACCREDITED MEASUREMENT CAPABILITIES

Sound Scope Matrix (CMC)

Calibration & Measurement Capabilities (CMC) accredited under ISO/IEC 17025:2017. Uncertainties expressed at an expanded confidence level of approx. 95.45% (k=2).

REQUEST OFFICIAL NABL CERTIFICATE SCOPE PDF
Discipline / MeasurandCalibration RangeTest ConditionExpanded CMC (±)Master Reference StandardTest Standard / Method
Sound Level Meters (Class 1 & Class 2)30 dB to 140 dB (Frequency: 31.5 Hz to 16 kHz)A, C, Z Frequency Weighting & Fast/Slow Time Weighting± 0.20 dBB&K Multi-Frequency Sound Calibrator + Audio Signal GenAcoustic & electrical testing as per IEC 61672-3:2013
Acoustic Calibrators & Pistonphones94 dB, 104 dB, 114 dB, 124 dBCoupler Cavity with Barometric Correction± 0.12 dBBrüel & Kjær Type 4228 Pistonphone + Digital BarometerSound calibrator performance verification (IEC 60942:2018)
COMPREHENSIVE EQUIPMENT DIRECTORY

Instruments Serviced & Calibrated

We calibrate all major industrial, pharmaceutical, aerospace, and research-grade instrumentation.

Class 1 & 2 Sound Level Meters

Brüel & Kjær, Rion, Testo, Casella, Cirrus SLMs

Personal Noise Dosimeters

Occupational health & industrial hygiene noise badges

Acoustic Sound Calibrators

94 dB / 114 dB portable verification calibrators

Condenser Measurement Microphones

1/2 inch and 1/4 inch free-field & pressure mics

METROLOGICAL INTEGRITY & RIGOR

Climate & Environment
Designed for Zero Drift

Every measurement at MSIR INDIA is conducted under strictly monitored and auditable environmental controls adhering to ISO/IEC 17025:2017:

Full floating-room anechoic chamber isolated from building structural acoustics
Real-time barometric pressure compensation applied to acoustic pistonphone output
A, C, and Z weighting curve filter response measured point-by-point across 20 Hz - 20 kHz
Ultra-low electrical noise amplifiers preventing 50 Hz hum injection into microphone signals

UNBROKEN CHAIN OF TRACEABILITY (ISO/IEC 17025)

TIER 1 // INTERNATIONAL DEFINITION

BIPM SI Pressure (Pascal) and Frequency (Hertz)

SI BASE
TIER 2 // NATIONAL METROLOGY INSTITUTE

NPL India Acoustic Metrology Primary Laboratory

PRIMARY LAB
TIER 3 // MSIR INDIA ACCREDITED LAB

MSIR Master B&K Pistonphone & Condenser Microphones (NABL CC-2891)

NABL CC-2891
TIER 4 // CLIENT TEST & MEASUREMENT ASSET

Client Sound Level Meters, Noise Dosimeters, and Microphones

CALIBRATED
Complies fully with ILAC P10 & ISO/IEC 17025:2017 traceability policy.
INSTANT ESTIMATE & INQUIRY

Book Sound Calibration

Get an instant fee estimate for your calibration scope or book an immediate laboratory slot with our certified metrologists.

Estimated Turnaround:2 Business Days
Calibration Certificate:NABL ISO/IEC 17025
Traceability:NPL India (Dual Signatory)

INDICATIVE CALIBRATION FEE

3,000 + GST

CONFIRM & SCHEDULE INTAKE
+91 94440 28911 lab@msirindia.com

Explore All 12 Metrology Laboratories

MSIR INDIA operates 12 specialized NABL-accredited metrology domains in Chennai.

RETURN TO LABORATORY DOMAINS