HomeMetrology DomainsElectrotechnical Facility
CHAMBER: 23.0 °C ± 0.5 °C
HUMIDITY: 45.0% RH
FARADAY CAGE: ACTIVE
NABL ISO/IEC 17025 ACCREDITED Primary Calibration Discipline Traceable to NPL (India) & BIPM

ELECTROTECHNICAL
CALIBRATION FACILITY

MSIR INDIA houses Chennai’s premier electrical metrology laboratory. Engineered with vibration-isolated optical breadboards, low-noise Faradaic electromagnetic shielding, and automated IEEE-488 GPIB data acquisition, we deliver sub-picoampere to 100 kV high-voltage calibration with world-class measurement uncertainty (down to ±1.5 ppm).

VOLTAGE RANGE

10 nV to 100 kV

DC & AC to 70 kV RMS

CURRENT RANGE

10 pA to 1,000 A

Up to 10 kA via coils

BEST MEASUREMENT CMC

± 1.5 ppm

Fluke 5730A 8.5-Digit

TIMEBASE STABILITY

2 × 10⁻¹¹

Rubidium Atomic Standard

LABORATORY SIGNAL SIMULATOR & TELEMETRY STUDIO

Real-Time Waveform & Metrology Analyzer

CHANNEL 1 // TRACEABLE PRIMARY REFERENCE BUS
ACQUISITION: IEEE-488 GPIB (100 kS/s)

MEASURED READOUT

10.000002

EXPANDED UNCERTAINTY (k=2)

± 1.20 ppm

TRACEABILITY

NPL (India)

PARAMETER CONTROLS

Test Amplitude:10 V
Inject Thermal EMF / EMI Noise:
Waveform Stream:
ISO 17025 Compliant Measurement

Readings undergo automated 20-sample averaging and Type A/B standard uncertainty evaluation (GUM JCGM 100:2008).

PRIMARY INSTRUMENTATION ARSENAL

Primary Reference Standards

Our laboratory calibration capabilities are backed by industry-standard calibrators, reference multimeters, and atomic clocks directly calibrated by NPL India and BIPM-recognized National Metrology Institutes.

Primary Metrology StandardISO 17025

Fluke 5730A

Multifunction Reference Calibrator

The global gold standard in DC/LF electrical calibration. Features internal Artifact Calibration to maintain traceability to external standards with sub-ppm stability.

Technical Highlights:

DC Voltage:0 to 1100 V (±1.5 ppm)
AC Voltage:220 mV to 1100 V (10 Hz - 1.2 MHz)
Resistance:1 Ω to 100 MΩ in x1 & x1.9
DC/AC Current:Up to 2.2 A (120 A with 52120A)
NPL India / NIST Direct Calibration
Reference Transfer StandardISO 17025

Keysight 3458A

8.5-Digit High-Stability DMM

Industry benchmark for precision measurement and linearity. Offers 8.5-digit resolution, transfer accuracy of 0.1 ppm, and sub-nanovolt DC voltage sampling.

Technical Highlights:

Resolution:8.5 Digits (10 nV sensitivity)
Transfer Accuracy:0.1 ppm of reading
Sampling Speed:Up to 100k readings/sec
Linearity:±0.05 ppm of reading + 0.05 ppm range
Traceable to BIPM Josephson Junction Array
High-Current Master SourceISO 17025

Fluke 52120A

Transconductance Power Current Amplifier

Delivers stable, high-accuracy AC and DC current outputs up to 120 Amperes standalone, and up to 10,000 A with parallel-coupled toroidal coils for clamp meters.

Technical Highlights:

Output Current:Up to 120 A DC & AC RMS
Coil Multiplier:Up to 12,000 A-turns (25/50-turn)
Bandwidth:DC to 100 kHz high fidelity
Compliance Voltage:Up to 4.5 V RMS
Traceable to National Metrology Institutes (NPLI)
Primary Resistance TransferISO 17025

Guildline 9330 Series

Ultra-Stable Resistance Standards in Oil Bath

Hermetically sealed standard resistors immersed in a temperature-regulated fluid bath maintained at 23.00 °C, ensuring virtually zero thermal drift.

Technical Highlights:

Resistance Values:0.001 Ω to 100 MΩ decade steps
Temp Coefficient:< 0.1 ppm / °C
Annual Stability:< 0.5 ppm / year
Bath Temperature:23.00 °C ± 0.01 °C regulated
Quantum Hall Resistance Traceable
Master Time & Frequency ReferenceISO 17025

SRS FS725

Rubidium Atomic Frequency Standard

Ultra-low phase noise rubidium atomic clock source providing a rock-solid 10 MHz reference for all frequency counters, signal generators, and spectrum analyzers.

Technical Highlights:

Outputs:10 MHz & 5 MHz Sine, 1 PPS
Allan Deviation:< 2 × 10⁻¹¹ (1 s)
Monthly Aging:< 5 × 10⁻¹¹ / month
Harmonic Distortion:< -30 dBc
Directly synchronized with UTC (NPLI)
High Voltage Breakdown FacilityISO 17025

Phenix Technologies 100kV

Precision AC/DC High Voltage Divider

Shielded high-voltage capacitive-resistive divider designed for ultra-high accuracy measurements of hipot test sets, megohmmeters, and spark testers.

Technical Highlights:

DC Voltage Range:0 to 100 kV DC
AC Voltage Range:0 to 70 kV AC RMS (50/60 Hz)
Voltage Ratio:10,000 : 1 precision matched
Measurement Uncertainty:± 0.05 % of reading
High Voltage Metrology NPL Traceable
ACCREDITED MEASUREMENT CAPABILITIES

Electrotechnical Scope Matrix (CMC)

Calibration & Measurement Capabilities (CMC) accredited under ISO/IEC 17025:2017. Uncertainties expressed at an expanded confidence level of approx. 95.45% (k=2).

REQUEST OFFICIAL NABL CERTIFICATE SCOPE PDF
Discipline / MeasurandCalibration RangeFrequency / Test ConditionExpanded CMC (±)Master Reference StandardTest Standard / Method
DC Voltage (Low Range / Sub-Microvolt)10 nV to 200 mVDC / Direct± (1.2 ppm to 3.5 ppm)Fluke 5730A Calibrator / Keysight 3458A DMMDirect measurement / Comparison as per EURAMET cg-15
DC Voltage (Standard & High Range)200 mV to 1100 VDC / Direct± (1.5 ppm to 4.0 ppm)Fluke 5730A Multifunction CalibratorDirect substitution against calibrated reference
DC High Voltage1.1 kV to 100 kVDC / Direct± 0.05 % to ± 0.12 %Phenix Precision HV Divider & 8.5-Digit DMMPrecision resistive divider ratio technique (IS 2071)
AC Voltage (Low & Audio Frequency)1 mV to 1000 V10 Hz to 100 kHz± 25 ppm to ± 90 ppmFluke 5730A with Wideband AC optionThermal voltage transfer comparison
AC Voltage (High Frequency / RF)10 mV to 10 V100 kHz to 30 MHz± 0.15 % to ± 0.45 %Precision Thermal Transfer StandardsRF Power & AC/DC thermal difference metrology
AC High Voltage (50 Hz / 60 Hz)1 kV to 70 kV RMS50 Hz / 60 Hz± 0.08 % of readingCalibrated Capacitive-Resistive DividerHigh voltage AC test standard (IEC 60060-2)
DC Current (Picoampere to Milliampere)10 pA to 200 mADC / Direct± 10 ppm to ± 35 ppmElectrometer Reference / Keysight 3458AStandard resistor IV conversion technique
DC Current (High Current / Shunt Range)200 mA to 1000 ADC / Direct± 45 ppm to ± 120 ppmFluke 52120A Amplifier & Precision Coaxial ShuntsPrecision low-inductance shunt drop measurement
AC Current10 µA to 1000 A10 Hz to 10 kHz± 80 ppm to ± 250 ppmFluke 5730A + 52120A Power TransconductanceAC current shunt comparison & Rogowski calibration
Standard Resistance (4-Wire Kelvin)10 µΩ to 1 Ω (Low Ohm)DC / Direct± 2.5 ppm to ± 10 ppmGuildline Standard Resistors in Oil BathAutomated 4-Wire current reversal null comparison
Resistance (Standard Decade Range)1 Ω to 100 MΩDC / Direct± 1.8 ppm to ± 15 ppmGuildline 9330 Series & Fluke 5730ADirect ratio bridge substitution
High & Ultra-High Resistance (Insulation)100 MΩ to 10 TΩ (Teraohm)DC / Direct± 0.25 % to ± 1.5 %High-Voltage Guarded Decade Megohm StandardSub-picoampere constant voltage dual-bridge method
Time & Frequency Standard0.001 Hz to 26.5 GHzDC / Direct± 2 × 10⁻¹¹ (Rubidium)SRS FS725 Atomic Standard + Keysight CounterDirect phase comparison against UTC reference
Active & Reactive AC Power (1-Phase / 3-Phase)10 V - 600 V / 10 mA - 100 A (PF: -1 to +1)45 Hz to 400 Hz± 0.025 % of apparent powerRadian / Fluke Precision Power StandardDual-channel digital sampling wattmeter method
COMPREHENSIVE EQUIPMENT DIRECTORY

Instruments Serviced & Calibrated

We service and calibrate all major brands including Keysight, Fluke, Tektronix, Rohde & Schwarz, Hioki, Megger, Chroma, Yokogawa, and Keithley.

Digital Multimeters

3.5 to 8.5 Digits, Benchtop & Handheld

Oscilloscopes

DSO, MSO, Sampling up to 26.5 GHz

AC/DC Clamp Meters

Up to 10,000 A with Coils & Rogowski

Power Quality Analyzers

Harmonics, Flicker, 3-Phase Energy

Insulation & Megohm Testers

Up to 10 kV and 10 TΩ Resistance

Hipot / Dielectric Testers

AC & DC High Voltage Withstand

LCR & Impedance Meters

Capacitance, Inductance, ESR, Tan δ

Earth & Ground Testers

Soil Resistivity, 2/3/4-Wire Grounding

Frequency Counters & Timers

Crystal Oscillators, Time Interval

Signal & Waveform Generators

Function, Arbitrary, Pulse, RF Sources

Current Shunts & Transducers

Precision DC & AC Coaxial Shunts

Process & Loop Calibrators

4-20 mA, 0-10 V, RTD, Thermocouple

METROLOGICAL INTEGRITY & CLEAN ENVIRONMENT

Engineered For Zero
Thermal & Electromagnetic Drift

Electrical metrology at the ppm level demands strict isolation from temperature shifts, relative humidity swings, electromagnetic interference, and seismic floor vibration. MSIR INDIA’s electrotechnical facility operates inside a custom-constructed environment:

Continuous 24/7 HVAC Control: Temperature held strictly at 23.0 °C ± 0.5 °C with independent multi-stage chillers.
Faraday RF Enclosure: Copper-mesh shielded bays attenuating wireless EMI, cellular signals, and industrial noise.
Dedicated Metrology Earthing (<0.2 Ω): Isolated solid chemical ground rods isolated from the building power grid.
Pneumatic Optical Tables: Sub-micron vibration damping protecting delicate galvanometer balances and quartz oscillators.

UNBROKEN CHAIN OF TRACEABILITY (ISO/IEC 17025)

TIER 1 // INTERNATIONAL METROLOGY

BIPM / SI Base Units (Volt, Ampere, Ohm, Second)

SI DEFINITION
TIER 2 // NATIONAL METROLOGY INSTITUTE

National Physical Laboratory (NPL India) / NIST

PRIMARY LAB
TIER 3 // MSIR INDIA ACCREDITED FACILITY

Fluke 5730A, Keysight 3458A, Rubidium Standard

NABL CC-2891
TIER 4 // CLIENT TEST & MEASUREMENT ASSET

Your Industrial Multimeters, Hipots, Oscilloscopes & Analyzers

CALIBRATED
Complies fully with ILAC P10 & ISO/IEC 17025:2017 traceability policy.
INSTANT ESTIMATE & INQUIRY

Book Electrotechnical Calibration

Get an instant estimate for your calibration scope or book an immediate laboratory slot with our certified metrologists.

Estimated Turnaround:2 Business Days
Calibration Certificate:NABL ISO/IEC 17025
Traceability:NPL India (Dual Signatory)

INDICATIVE CALIBRATION FEE

3,500 + GST

CONFIRM & SCHEDULE INTAKE
+91 94440 28911 lab@msirindia.com

Explore Other Metrology Laboratories

MSIR INDIA houses 12 accredited calibration laboratories under one roof in Chennai.

RETURN TO ALL 12 LAB DOMAINS